Difference between revisions of "20200508-PREX2-Analysis"
From PREX Wiki
Jump to navigationJump to searchLine 13: | Line 13: | ||
== Agenda == | == Agenda == | ||
− | + | #* Caryn : Review Pedestal [http://ace.phys.virginia.edu/HAPPEX/4095 HAPLOG 4095] | |
− | [ | + | #** KP : is there any change in digital BCM? |
+ | #** CP : there is v2f connection change and chassis swap | ||
+ | #** KP : so are we using different pedestal for them ? | ||
+ | #** CP : choice of normalizing BCM, we will have digital BCM in a few runs and other uses analog BCM | ||
+ | #** CP : PREX changes PMT HV when we changed to 70 uA and a change of quartz bar. | ||
+ | #** KP : pedestal history shows difference between configuration. | ||
+ | #** CP : regarding error bar, everything uses rms/sqrt(N) except of unser-calibration which uses rms. | ||
+ | #** CP : for 85 uA calibration, only includes upper half data points for better linearity | ||
+ | #** KP : do we see non-linearity? | ||
+ | #** DA : we didnt' change HV for 85 uA period | ||
+ | #** CP : the analog BCM was moved to different ADC panels and needs extra pedestal | ||
=== Analysis projects === | === Analysis projects === | ||
Line 26: | Line 36: | ||
== Present == | == Present == | ||
− | Kent P, Adam Z, Cameron C, Caryn P, Chandan G, Ciprian G, David A, Devi A, Dmitrii N, Don J, Dustin M, Eric K, Ezekiel W, Hanjie L, Jay B, Juan Carlos C, KK, Paul K, Robert R, Ryan R, Ye T, Yufan C, Weibin Z, Victoria O, Siyu J, Silviu CD, Quinn C, Amali P, Sanghwa P, | + | Kent P, Adam Z, Cameron C, Caryn P, Chandan G, Ciprian G, David A, Devi A, Dmitrii N, Don J, Dustin M, Eric K, Ezekiel W, Hanjie L, Jay B, Juan Carlos C, KK, Paul K, Robert R, Ryan R, Ye T, Yufan C, Weibin Z, Victoria O, Siyu J, Silviu CD, Quinn C, Amali P, Sanghwa P, Paul S, Sakib R, Brian Q, Robert M |
===Excused=== | ===Excused=== | ||
Latest revision as of 13:58, 8 May 2020
Back to Main Page >> PREX2_Analysis
previous meeting << >> following meeting
Logistic information
Meeting room at JLab: REMOTE ONLY
BlueJeans calling instructions: Toll-Free Number (U.S.& Canada): 888-240-2560 International toll number: 408-740-7256 Bluejeans CODE: 206 308 0414 Bluejeans link: https://bluejeans.com/2063080414
Agenda
- Caryn : Review Pedestal HAPLOG 4095
- KP : is there any change in digital BCM?
- CP : there is v2f connection change and chassis swap
- KP : so are we using different pedestal for them ?
- CP : choice of normalizing BCM, we will have digital BCM in a few runs and other uses analog BCM
- CP : PREX changes PMT HV when we changed to 70 uA and a change of quartz bar.
- KP : pedestal history shows difference between configuration.
- CP : regarding error bar, everything uses rms/sqrt(N) except of unser-calibration which uses rms.
- CP : for 85 uA calibration, only includes upper half data points for better linearity
- KP : do we see non-linearity?
- DA : we didnt' change HV for 85 uA period
- CP : the analog BCM was moved to different ADC panels and needs extra pedestal
- Caryn : Review Pedestal HAPLOG 4095
Analysis projects
- BCM double difference study
- BPM resolution
- Pedestal / Normalization BCM history
- Events/ Patterns Counter
- Dithering sensitivity/Quality re-check
- Grand statistics re-check
- Linearity check
Present
Kent P, Adam Z, Cameron C, Caryn P, Chandan G, Ciprian G, David A, Devi A, Dmitrii N, Don J, Dustin M, Eric K, Ezekiel W, Hanjie L, Jay B, Juan Carlos C, KK, Paul K, Robert R, Ryan R, Ye T, Yufan C, Weibin Z, Victoria O, Siyu J, Silviu CD, Quinn C, Amali P, Sanghwa P, Paul S, Sakib R, Brian Q, Robert M