Back to [[Main_Page|Main Page]] >> [[DAQ_Doc_Portal|DAQ Documentation Portal]] >> [[DAQ_Testing|DAQ Testing]] >> [[DAQ_Commissioning_Notes|DAQ Commissioning Notes]] | Back to [[Main_Page|Main Page]] >> [[DAQ_Doc_Portal|DAQ Documentation Portal]] >> [[DAQ_Testing|DAQ Testing]] >> [[DAQ_Commissioning_Notes|DAQ Commissioning Notes]] | ||
− | [[DAQ_Testing/ | + | [[DAQ_Testing/20180731|Previous Day of Testing]] << >> [[DAQ_Testing/20181101|Next Day of Testing]] |
− | + | September 10th, 2018 Testers: Cameron | |
See [[DAQ_Meeting/20180810|Today's Meeting Notes]] | See [[DAQ_Meeting/20180810|Today's Meeting Notes]] |
Latest revision as of 19:47, 2 November 2018
Back to Main Page >> DAQ Documentation Portal >> DAQ Testing >> DAQ Commissioning Notes
Previous Day of Testing << >> Next Day of Testing
September 10th, 2018 Testers: Cameron
Goals
Replace battery in LHRS ROC 25 (hallavme14) so that NVRAM doesn't get erased after each power cycle.
Status
- The battery (small yellow rectangle with 4 pins) used as a replacement is labelled "990ST M4T28-BR12SH1 TIMEKEEPER SNAPHAT MYS 99 611" and Cameron has 4 more copies with the same label, plus the old one that was probably bad.
- Replacing the battery is easy, you just pull out the ROC and the battery is installed over a integrated chip next to the serial port on the bottom left and it just snaps into place.
- I did not do a power cycle test in the hall, but I did restore the RAM remotely (from apar@adaq3:~/doc/vmeram.doc)
- While moving the modules around the TI to Happex Timing Board twisted pair cable came undone and I did not put it back.
- This means that for now the LHRS can only run in Trigger Supervisor mode (not standalone HapTB mode).
To Do
- Verify the RAM stays by doing a hardware power cycle
- Replicate battery replacement in all other ROCs to prevent this issue from reappearing (Cameron has the batteries)
- Replace the HAPTB DAC to TI module twisted pair connection
- Add HAPPEX Timing Board to Injector
- Test all ADCs in Injector
- Make a bank for all ADCs