Difference between revisions of "Target density variation check"

From PREX Wiki
Jump to navigationJump to search
Line 12: Line 12:
 
** [https://logbooks.jlab.org/entry/3727983 HALOG summary]
 
** [https://logbooks.jlab.org/entry/3727983 HALOG summary]
 
** Took high statistics run at very low current (30nA requested), Target #10
 
** Took high statistics run at very low current (30nA requested), Target #10
 +
** Septum at nominal setting, no Sieve
 
** Similar result shown in RHRS data as previous high current test
 
** Similar result shown in RHRS data as previous high current test
 
** Different distribution in LHRS is puzzling
 
** Different distribution in LHRS is puzzling

Revision as of 15:21, 3 September 2019

  • First measurement 09/02/2019
    • HALOG summary
    • Target #7 and #10 were checked for current range 5-70uA
    • Septum mistuned on purpose in order to reduce rates, T2 trigger is used
    • Sieve in
    • Clear non-uniformity observed from degraded target (#10) compared to good (still in use) target (#7).
    • More visible at higher current, but hint of similar signature of density rearrangement seen at low current.
    • Consistent results from both LHRS and RHRS spot data


  • Second measurement 09/03/2019
    • HALOG summary
    • Took high statistics run at very low current (30nA requested), Target #10
    • Septum at nominal setting, no Sieve
    • Similar result shown in RHRS data as previous high current test
    • Different distribution in LHRS is puzzling