Difference between revisions of "CREXRunPlan-August20"
From PREX Wiki
Jump to navigationJump to searchKent Paschke (talk | contribs) |
Kent Paschke (talk | contribs) |
||
(13 intermediate revisions by 2 users not shown) | |||
Line 3: | Line 3: | ||
− | == CREX August 2020 Run Plan for Commissioning and Auxiliary Measurements | + | == CREX August 2020 Run Plan for Commissioning and Auxiliary Measurements (In Development) == |
− | See also the [[RunPlanDaily_CREX| Daily Run Plan]] and [[Run plan history CREX| Previous Daily Run Plans]] and [[CREXAuxiliaryMeasurements| Auxiliary Measurements ]] | + | See also the [[RunPlanDaily_CREX| Daily Run Plan]] and [[Run plan history CREX| Previous Daily Run Plans]] and [[CREXAuxiliaryMeasurements-Aug2020| Fall 2020 Auxiliary Measurements ]] |
[[PREX_I_Run_Plans | PREX I - 2010 Run Plan]] | [[PREX_I_Run_Plans | PREX I - 2010 Run Plan]] | ||
Line 11: | Line 11: | ||
[[PREX2AuxiliaryMeasurements| PREX2 2019 Auxiliary Measurements ]] | [[PREX2AuxiliaryMeasurements| PREX2 2019 Auxiliary Measurements ]] | ||
− | + | Original [[CREXRunPlan | CREX Run Plan]] and original [[CREXAuxiliaryMeasurements| CREX Auxiliary Measurements ]] | |
− | + | == Beam Restoration and Low Current Commissioning == | |
+ | # Basic beam setup | ||
+ | #* establish tune beam and low current (up to 2uA?) CW beam in the hall. <font color="blue"> --Done </font> | ||
+ | ## Establish beam to dump. <font color="blue"> --Done </font> | ||
+ | ## dump ion chamber calibration? (Low current.) <font color="blue"> --Done </font> | ||
+ | ##* assume collimator centering as before, trajectory as before | ||
+ | ##* spectrometer magnets on <font color="blue"> --Done </font> | ||
+ | ##* insert optics C-hole target, verify target location and raster calibration with Spot++ <font color="blue"> --Done </font> | ||
+ | ##** Presumably also need low current ion calibration for C-hole target? <font color="blue"> --Done </font> | ||
+ | ##* If hole is significantly different, re-verify BPM positions, log present spot++ and link comparison spot++. Call Kent (434-964-6777). | ||
+ | # First Spectrometer optics / chamber calibration check | ||
+ | #* Carbon optics target, sieve in (verify acceptance, verify optics db with new VDC alignment/calibration). <font color="blue"> --Done </font> | ||
+ | #* same, with GEM data <font color="blue"> --Done </font> | ||
+ | #* Verify Cold target position | ||
+ | ## spot++ on C-hole (low current IC calibration?) <font color="blue"> --Done </font> | ||
+ | ## spot++ on Ca-48 (low current IC calibration?) <font color="blue"> --Done </font> | ||
+ | # Detector alignment <font color="blue"> --Defer for later </font> | ||
+ | ## start with initial detector guess for detector alignment. (sieve out, Ca-48 target) | ||
+ | ##* Do not refine for small details! +/- 2mm is fine) | ||
+ | ##* Do not adjust At detector alignment at this time | ||
+ | ## if right arm not functional, use GEMs with Chandan's GEM alignment? | ||
+ | ## Take Q2 data (longer run after alignment selected) | ||
+ | # Optics: | ||
+ | ## C-foil optics data (sieve in, dp=+1,0,-1) <font color="blue"> --done dp=-1,0; defer - need dp=1 with sieve, and sieve out C-foil data</font> | ||
+ | ## water cell data (sieve in dp=+1,0,-1) <font color="blue"> --done dp=-1,0; defer - need dp=1 with sieve and sieve out watercell data </font> | ||
+ | ## Do we want: | ||
+ | ##* repeat harp/bullseye scan? <font color="blue"> --Done </font> | ||
+ | ##* repeat collimator centering scan? <font color="blue"> --may be we can skip this </font> | ||
− | <font color=" | + | == High current checkout == |
+ | |||
+ | # Establish high current to hall dump. 150uA. Radiation, IC levels should match expectation. | ||
+ | # <font color="blue">Ion Chamber calibration - thick Target </font> <span style="background-color:#F1C40F">0.25 shift</span> | ||
+ | ## Should be easy/standard procedure, but verify calibration matches old calibration <font color="blue"> --Done </font> | ||
+ | ## verify trip points match old trip points, running conditions for high current match old high current running. If not, differences must be understood. <font color="blue"> --Matched pretty good </font> | ||
+ | # Setup beam through Compton chicane. <font color="blue"> --pnding </font> | ||
+ | #* must keep watch on Compton rates whenever through Compton chicane <font color="blue"> --ongoing </font> | ||
+ | # FFB commissioning. Expected to be used when running high current <font color="blue"> --ongoing </font> | ||
+ | # Establish beam on cold target <font color="blue"> --Done </font> | ||
+ | #* integrating mode detectors <font color="blue"> --Done </font> | ||
+ | ## C-hole, spot++ | ||
+ | ## Thick carbon target. Spot++ to verify on target | ||
+ | ## Thick carbon to high current. | ||
+ | ## Ca-48 target. Spot++ to verify on target | ||
+ | ## Ca-48 to high current. Burn off (30 minutes each at 50,100,150uA) <font color="blue"> Call Silviu just before starting this </font> | ||
+ | # Verify beam modulation functionality <font color="blue"> --Done </font> | ||
+ | #* low current test, then high current to watch of Ion Chamber rates or beam loss | ||
+ | #* Ignore orthogonality for first test | ||
+ | # Test production with 1-2 hours. <font color="blue"> --Done </font> | ||
+ | ## detector and SAM widths | ||
+ | ## regression sensitivities | ||
+ | ## beam noise and asymmetry check | ||
+ | ## collimator water and thermocouple temperatures, | ||
+ | ## radiation monitors and ion chamber stability | ||
+ | ## Tune beam modulation orthogonality, if needed, after 1-2 hours, then take at least 1 hour data to confirm test. | ||
− | + | == <font color="magenta"> '''Start Production''' </font> == | |
− | == Beam | + | == Other Early Tests and Checks (first few days) == |
− | # <font color="blue"> | + | # Beam Monitor calibration |
− | # | + | #* soon after restart |
− | ## | + | ## Run current scan using Unser to calibrate beam charge monitor readout in parity DAQ. |
+ | ##* Run the Compton DAQ concurrently, for calibration | ||
+ | ## Go to FixedGain when at 150 uA on bpms (list of crates???) and repeat local current scan to set bpm pedestals. | ||
+ | # Fine-tune detector alignment | ||
+ | # At detector alignment | ||
+ | # Compton commissioning | ||
+ | # Moller measurement | ||
+ | |||
+ | == Other Tests and Checks == | ||
+ | # Do we need to repeat spin-dance? | ||
+ | # <font color="blue">Linearity Studies</font> | ||
+ | # Repeat Q2 and pointing measurements | ||
+ | # Arc Energy Measurement | ||
+ | # <font color="blue">Background Studies</font> | ||
+ | ## Scans of Septum Magnet and HRS Dipole | ||
+ | # Possible scan of Q1 to optimize acceptance and verify that collimator defines acceptance. (old idea, still relevant?) | ||
+ | |||
+ | == Polarimetry Studies == | ||
+ | |||
+ | # <font color="blue">Compton Commissioning</font> | ||
+ | ## Beam Tune, Background reduction | ||
+ | ## Compton Cavity Checkout | ||
+ | ### Establish laser in cavity | ||
+ | ### Move photon detector to visual center, scan beam for collisions | ||
+ | ### Establish orbit lock | ||
+ | ## Photon Detector Checkout | ||
+ | ### Detector power-up, find optimal gain. | ||
+ | ### Center detector with fingers. | ||
+ | ### 15 minute spectrum and rate study. | ||
+ | ### Brief target-out run? (Probably not necessary but may be desirable) | ||
+ | ### Polarization commissioning run (up to 2 hour target-in) | ||
+ | ### Jaws checkout (?) | ||
+ | # <font color="blue">Moller Commissioning</font> -- Simona Malace, Don Jones, et.al. | ||
+ | |||
+ | |||
+ | |||
+ | <!--- | ||
## Design match to hall expected as part of original configuration, if additional beam time is required will be necessary to coordinate with optics/production runplan | ## Design match to hall expected as part of original configuration, if additional beam time is required will be necessary to coordinate with optics/production runplan | ||
## ion chamber calibration (should be ready for this if not already done). We should have targets ready; neeed targets in (MASK ION CHAMBERS); <font color="red"> Dustin will coordinate</font> | ## ion chamber calibration (should be ready for this if not already done). We should have targets ready; neeed targets in (MASK ION CHAMBERS); <font color="red"> Dustin will coordinate</font> | ||
Line 30: | Line 118: | ||
## Characterize injector optics configuration <font color="green"> done </font> | ## Characterize injector optics configuration <font color="green"> done </font> | ||
+ | ---> | ||
+ | |||
+ | <!-- | ||
== Low Current Commissioning == | == Low Current Commissioning == | ||
* Basic beamline commissioning (may be 3 shifts total) | * Basic beamline commissioning (may be 3 shifts total) | ||
Line 81: | Line 172: | ||
## Move to Ca-40 target, location #(?) | ## Move to Ca-40 target, location #(?) | ||
## establish 5uA CW, 2x2 raster. Run spot++ to verify raster size, not hitting edges. Log it. | ## establish 5uA CW, 2x2 raster. Run spot++ to verify raster size, not hitting edges. Log it. | ||
− | # <font color="blue"> Very low current commissioning </font><span style="background-color:#F1C40F"></span> '''Very low current not needed because 1 uA is only few kHz in detectors.''' <!-- | + | # <font color="blue"> Very low current commissioning </font><span style="background-color:#F1C40F"></span> '''Very low current not needed because 1 uA is only few kHz in detectors.''' |
+ | --> | ||
+ | |||
+ | <!-- | ||
'''LOW CURRENT CAVITIES WILL BE SET UP THE SAME AS FOR PREX-II (so don't need this)''' | '''LOW CURRENT CAVITIES WILL BE SET UP THE SAME AS FOR PREX-II (so don't need this)''' | ||
#* commission low current cavities relative to stripline calibration. Commission cavity bpm lock. | #* commission low current cavities relative to stripline calibration. Commission cavity bpm lock. | ||
Line 102: | Line 196: | ||
####Look at the strip chart of IPM1H04BX, IPM1H04CX, IPM1H04DX and IPM1H04BY, IPM1H04CY, (current signal) IPM1H04DY, IPM1H04B, IPM1H04C, IPM1H04D, if the signal is saturate or too small to present. Then we need to change the gain of the cavity BPMs | ####Look at the strip chart of IPM1H04BX, IPM1H04CX, IPM1H04DX and IPM1H04BY, IPM1H04CY, (current signal) IPM1H04DY, IPM1H04B, IPM1H04C, IPM1H04D, if the signal is saturate or too small to present. Then we need to change the gain of the cavity BPMs | ||
#####Change the gain, please call Ye Tian (803-553-3570) --> | #####Change the gain, please call Ye Tian (803-553-3570) --> | ||
+ | |||
+ | <!-- | ||
== Spectrometer Commissioning == | == Spectrometer Commissioning == | ||
Line 148: | Line 244: | ||
##* Look at septum hole pattern, check no missing holes (if sieve is in) | ##* Look at septum hole pattern, check no missing holes (if sieve is in) | ||
## Check eigenvector quads steps of +/- 1% on Q1? | ## Check eigenvector quads steps of +/- 1% on Q1? | ||
− | ##* Analyze for Left/Right x and y width. | + | ##* Analyze for Left/Right x and y width. |
− | |||
## Expected: width should be minimized | ## Expected: width should be minimized | ||
##* Tune that minimizes width at detectors is new central tune. | ##* Tune that minimizes width at detectors is new central tune. | ||
Line 201: | Line 296: | ||
# <font color="blue"> Position At detectors </font>: Dustin, Ryan | # <font color="blue"> Position At detectors </font>: Dustin, Ryan | ||
− | + | --> | |
− | |||
− | + | <!-- | |
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
− | |||
[[CREX Detailed High current plan|detailed plan]] | [[CREX Detailed High current plan|detailed plan]] | ||
Line 319: | Line 405: | ||
== Important Activities After First Few Days == | == Important Activities After First Few Days == | ||
− | # <font color="blue">Linearity Studies</font> | + | # <font color="blue">Linearity Studies</font> |
− | # <font color="blue">Background Studies</font> | + | # <font color="blue">Background Studies</font> |
## Scans of Septum Magnet and HRS Dipole | ## Scans of Septum Magnet and HRS Dipole | ||
## Thin Lead Target to check for inelastics at high-resolution | ## Thin Lead Target to check for inelastics at high-resolution | ||
Line 353: | Line 439: | ||
# Repeat Q2 and pointing measurements | # Repeat Q2 and pointing measurements | ||
# Arc Energy Measurement | # Arc Energy Measurement | ||
+ | |||
+ | --> |
Latest revision as of 20:28, 11 September 2020
CREX August 2020 Run Plan for Commissioning and Auxiliary Measurements (In Development)
See also the Daily Run Plan and Previous Daily Run Plans and Fall 2020 Auxiliary Measurements
PREX I - 2010 Run Plan PREX II - 2019 Run Plan PREX2 2019 Auxiliary Measurements
Original CREX Run Plan and original CREX Auxiliary Measurements
Beam Restoration and Low Current Commissioning
- Basic beam setup
- establish tune beam and low current (up to 2uA?) CW beam in the hall. --Done
- Establish beam to dump. --Done
- dump ion chamber calibration? (Low current.) --Done
- assume collimator centering as before, trajectory as before
- spectrometer magnets on --Done
- insert optics C-hole target, verify target location and raster calibration with Spot++ --Done
- Presumably also need low current ion calibration for C-hole target? --Done
- If hole is significantly different, re-verify BPM positions, log present spot++ and link comparison spot++. Call Kent (434-964-6777).
- First Spectrometer optics / chamber calibration check
- Carbon optics target, sieve in (verify acceptance, verify optics db with new VDC alignment/calibration). --Done
- same, with GEM data --Done
- Verify Cold target position
- spot++ on C-hole (low current IC calibration?) --Done
- spot++ on Ca-48 (low current IC calibration?) --Done
- Detector alignment --Defer for later
- start with initial detector guess for detector alignment. (sieve out, Ca-48 target)
- Do not refine for small details! +/- 2mm is fine)
- Do not adjust At detector alignment at this time
- if right arm not functional, use GEMs with Chandan's GEM alignment?
- Take Q2 data (longer run after alignment selected)
- start with initial detector guess for detector alignment. (sieve out, Ca-48 target)
- Optics:
- C-foil optics data (sieve in, dp=+1,0,-1) --done dp=-1,0; defer - need dp=1 with sieve, and sieve out C-foil data
- water cell data (sieve in dp=+1,0,-1) --done dp=-1,0; defer - need dp=1 with sieve and sieve out watercell data
- Do we want:
- repeat harp/bullseye scan? --Done
- repeat collimator centering scan? --may be we can skip this
High current checkout
- Establish high current to hall dump. 150uA. Radiation, IC levels should match expectation.
- Ion Chamber calibration - thick Target 0.25 shift
- Should be easy/standard procedure, but verify calibration matches old calibration --Done
- verify trip points match old trip points, running conditions for high current match old high current running. If not, differences must be understood. --Matched pretty good
- Setup beam through Compton chicane. --pnding
- must keep watch on Compton rates whenever through Compton chicane --ongoing
- FFB commissioning. Expected to be used when running high current --ongoing
- Establish beam on cold target --Done
- integrating mode detectors --Done
- C-hole, spot++
- Thick carbon target. Spot++ to verify on target
- Thick carbon to high current.
- Ca-48 target. Spot++ to verify on target
- Ca-48 to high current. Burn off (30 minutes each at 50,100,150uA) Call Silviu just before starting this
- Verify beam modulation functionality --Done
- low current test, then high current to watch of Ion Chamber rates or beam loss
- Ignore orthogonality for first test
- Test production with 1-2 hours. --Done
- detector and SAM widths
- regression sensitivities
- beam noise and asymmetry check
- collimator water and thermocouple temperatures,
- radiation monitors and ion chamber stability
- Tune beam modulation orthogonality, if needed, after 1-2 hours, then take at least 1 hour data to confirm test.
Start Production
Other Early Tests and Checks (first few days)
- Beam Monitor calibration
- soon after restart
- Run current scan using Unser to calibrate beam charge monitor readout in parity DAQ.
- Run the Compton DAQ concurrently, for calibration
- Go to FixedGain when at 150 uA on bpms (list of crates???) and repeat local current scan to set bpm pedestals.
- Fine-tune detector alignment
- At detector alignment
- Compton commissioning
- Moller measurement
Other Tests and Checks
- Do we need to repeat spin-dance?
- Linearity Studies
- Repeat Q2 and pointing measurements
- Arc Energy Measurement
- Background Studies
- Scans of Septum Magnet and HRS Dipole
- Possible scan of Q1 to optimize acceptance and verify that collimator defines acceptance. (old idea, still relevant?)
Polarimetry Studies
- Compton Commissioning
- Beam Tune, Background reduction
- Compton Cavity Checkout
- Establish laser in cavity
- Move photon detector to visual center, scan beam for collisions
- Establish orbit lock
- Photon Detector Checkout
- Detector power-up, find optimal gain.
- Center detector with fingers.
- 15 minute spectrum and rate study.
- Brief target-out run? (Probably not necessary but may be desirable)
- Polarization commissioning run (up to 2 hour target-in)
- Jaws checkout (?)
- Moller Commissioning -- Simona Malace, Don Jones, et.al.