Back to [[Main_Page|Main Page]] >> [[DAQ_Doc_Portal|DAQ Documentation Portal]] >> [[DAQ_Testing|DAQ Testing]] >> [[DAQ_Commissioning_Notes|DAQ Commissioning Notes]] | Back to [[Main_Page|Main Page]] >> [[DAQ_Doc_Portal|DAQ Documentation Portal]] >> [[DAQ_Testing|DAQ Testing]] >> [[DAQ_Commissioning_Notes|DAQ Commissioning Notes]] | ||
− | [[DAQ_Testing/20180730|Previous Day of Testing]] << >> [[DAQ_Testing/ | + | [[DAQ_Testing/20180730|Previous Day of Testing]] << >> [[DAQ_Testing/20180910|Next Day of Testing]] |
− | July | + | July 31st, 2018 Testers: Cameron and Chandan |
See [[DAQ_Meeting/20180731|Today's Meeting Notes]] | See [[DAQ_Meeting/20180731|Today's Meeting Notes]] |
Latest revision as of 19:47, 2 November 2018
Back to Main Page >> DAQ Documentation Portal >> DAQ Testing >> DAQ Commissioning Notes
Previous Day of Testing << >> Next Day of Testing
July 31st, 2018 Testers: Cameron and Chandan
Goals
Test Injector NIM crate voltages to see if the error in there was a module or crate issue
Testing Injector Parity DAQ NIM logic crate
Using our voltmeter we saw that the -6V channel was off
- It read -4.2 volts instead of -6V
- So the issue was likely with the crate, not the old module
Result
- We know that the issue from July 27th was likely caused by a bad NIM crate, not the module itself
To Do
- Replace the NIM crate
- Rewire everything in the Injector DAQ
- Label everything
- Write out a diagram of the logic and the channels