* Run 7536 - https://logbooks.jlab.org/entry/3820478
+
* Run 7535 - Show the failure of ADC 032 channel 02, railed at -10 (as Victoria found a few days ago)
 +
* Run 7536 - Show the old LHRS ADC 025 with non-linear response: https://logbooks.jlab.org/entry/3820478
  
 
== Checking out all ADC modules ==
 
== Checking out all ADC modules ==
Line 21: Line 22:
 
* Run 7547-7548 - RHRS tests, both ADCs good... SAM 5 has some 60 Hz noise that is affected by jostling cables... maybe removed now after replugging
 
* Run 7547-7548 - RHRS tests, both ADCs good... SAM 5 has some 60 Hz noise that is affected by jostling cables... maybe removed now after replugging
 
* Run 7549 - LHRS test again, with ADC 001 this time, all good
 
* Run 7549 - LHRS test again, with ADC 001 this time, all good
 +
 +
HALOG entry: https://logbooks.jlab.org/entry/3820580#
  
 
[[Category:DAQ_Testing]]
 
[[Category:DAQ_Testing]]

Latest revision as of 17:04, 4 August 2020

Back to Main Page >> DAQ Documentation Portal >> DAQ Testing >> DAQ Commissioning Notes

Previous Day of Testing << >> Next Day of Testing

August 4th, 2020 Tester: Cameron Clarke

Replacing VQWK number 025 from L:HRS _0_3 bad channel

To make these plots I have used the DAQ_Checkout.conf decoding configuration (for the CH_INJ CODA config) and used the panguin macro QualityAssurance/testChannels.cfg

  • Run 7535 - Show the failure of ADC 032 channel 02, railed at -10 (as Victoria found a few days ago)
  • Run 7536 - Show the old LHRS ADC 025 with non-linear response: https://logbooks.jlab.org/entry/3820478

Checking out all ADC modules

Using the RampAll config (rampdac16 setting in CH crate cedit config)

  • Run 7537 - Looks good by eye, if interested see PDFs in ~/cameronc/DAQ-calibration/CREX-August-Restart/
  • Run 7538-7544 : remaining 7 ADCs in CH crate, all appear good
  • Run 7545 - added rampdac16 to L and RHRS in CEDIT - test run, works
  • Run 7546 - LHRS test, turns out channel _3 has another problem, no negative voltage range and strange positive channels-volts scale factor -> Will replace with ADC 001
  • Run 7547-7548 - RHRS tests, both ADCs good... SAM 5 has some 60 Hz noise that is affected by jostling cables... maybe removed now after replugging
  • Run 7549 - LHRS test again, with ADC 001 this time, all good

HALOG entry: https://logbooks.jlab.org/entry/3820580#