Difference between revisions of "CREXRunPlan-August20"
From PREX Wiki
Jump to navigationJump to searchKent Paschke (talk | contribs) |
Chandabindu (talk | contribs) |
||
Line 15: | Line 15: | ||
== Beam Restoration and Low Current Commissioning == | == Beam Restoration and Low Current Commissioning == | ||
# Basic beam setup | # Basic beam setup | ||
− | #* establish tune beam and low current (up to 2uA?) CW beam in the hall. | + | #* establish tune beam and low current (up to 2uA?) CW beam in the hall. <font color="blue"> --Done </font> |
− | ## Establish beam to dump. | + | ## Establish beam to dump. <font color="blue"> --Done </font> |
− | ## dump ion chamber calibration? (Low current.) | + | ## dump ion chamber calibration? (Low current.) <font color="blue"> --Done </font> |
##* assume collimator centering as before, trajectory as before | ##* assume collimator centering as before, trajectory as before | ||
− | ##* spectrometer magnets on | + | ##* spectrometer magnets on <font color="blue"> --Done </font> |
− | ##* insert optics C-hole target, verify target location and raster calibration with Spot++ | + | ##* insert optics C-hole target, verify target location and raster calibration with Spot++ <font color="blue"> --Done </font> |
− | ##** Presumably also need low current ion calibration for C-hole target? | + | ##** Presumably also need low current ion calibration for C-hole target? <font color="blue"> --Done </font> |
##* If hole is significantly different, re-verify BPM positions, log present spot++ and link comparison spot++. Call Kent (434-964-6777). | ##* If hole is significantly different, re-verify BPM positions, log present spot++ and link comparison spot++. Call Kent (434-964-6777). | ||
# First Spectrometer optics / chamber calibration check | # First Spectrometer optics / chamber calibration check | ||
− | #* Carbon optics target, sieve in (verify acceptance, verify optics db with new VDC alignment/calibration). | + | #* Carbon optics target, sieve in (verify acceptance, verify optics db with new VDC alignment/calibration). <font color="blue"> --Done </font> |
− | #* same, with GEM data | + | #* same, with GEM data <font color="blue"> --Done </font> |
#* Verify Cold target position | #* Verify Cold target position | ||
− | ## spot++ on C-hole (low current IC calibration?) | + | ## spot++ on C-hole (low current IC calibration?) <font color="blue"> --Done </font> |
− | ## spot++ on Ca-48 (low current IC calibration?) | + | ## spot++ on Ca-48 (low current IC calibration?) <font color="blue"> --Done </font> |
# Detector alignment | # Detector alignment | ||
## start with initial detector guess for detector alignment. (sieve out, Ca-48 target) | ## start with initial detector guess for detector alignment. (sieve out, Ca-48 target) | ||
Line 35: | Line 35: | ||
## if right arm not functional, use GEMs with Chandan's GEM alignment? | ## if right arm not functional, use GEMs with Chandan's GEM alignment? | ||
## Take Q2 data (longer run after alignment selected) | ## Take Q2 data (longer run after alignment selected) | ||
− | ## C-foil optics data (sieve in, dp=+1,0,-1) | + | ## C-foil optics data (sieve in, dp=+1,0,-1) <font color="blue"> --ongoing </font> |
# Deferred to later: | # Deferred to later: | ||
− | ## water cell data (sieve in dp=+1,0,-1) | + | ## water cell data (sieve in dp=+1,0,-1) <font color="blue"> --ongoing </font> |
## At detector alignment | ## At detector alignment | ||
## Do we want: | ## Do we want: | ||
− | ##* repeat harp/bullseye scan? | + | ##* repeat harp/bullseye scan? <font color="blue"> --Planned </font> |
− | ##* repeat collimator centering scan? | + | ##* repeat collimator centering scan? <font color="blue"> --may be we can skip this </font> |
== High current checkout == | == High current checkout == |
Revision as of 10:14, 9 August 2020
CREX August 2020 Run Plan for Commissioning and Auxiliary Measurements (In Development)
See also the Daily Run Plan and Previous Daily Run Plans and Auxiliary Measurements
PREX I - 2010 Run Plan PREX II - 2019 Run Plan PREX2 2019 Auxiliary Measurements
Original CREX Run Plan
Beam Restoration and Low Current Commissioning
- Basic beam setup
- establish tune beam and low current (up to 2uA?) CW beam in the hall. --Done
- Establish beam to dump. --Done
- dump ion chamber calibration? (Low current.) --Done
- assume collimator centering as before, trajectory as before
- spectrometer magnets on --Done
- insert optics C-hole target, verify target location and raster calibration with Spot++ --Done
- Presumably also need low current ion calibration for C-hole target? --Done
- If hole is significantly different, re-verify BPM positions, log present spot++ and link comparison spot++. Call Kent (434-964-6777).
- First Spectrometer optics / chamber calibration check
- Carbon optics target, sieve in (verify acceptance, verify optics db with new VDC alignment/calibration). --Done
- same, with GEM data --Done
- Verify Cold target position
- spot++ on C-hole (low current IC calibration?) --Done
- spot++ on Ca-48 (low current IC calibration?) --Done
- Detector alignment
- start with initial detector guess for detector alignment. (sieve out, Ca-48 target)
- Do not refine for small details! +/- 2mm is fine)
- Do not adjust At detector alignment at this time
- if right arm not functional, use GEMs with Chandan's GEM alignment?
- Take Q2 data (longer run after alignment selected)
- C-foil optics data (sieve in, dp=+1,0,-1) --ongoing
- start with initial detector guess for detector alignment. (sieve out, Ca-48 target)
- Deferred to later:
- water cell data (sieve in dp=+1,0,-1) --ongoing
- At detector alignment
- Do we want:
- repeat harp/bullseye scan? --Planned
- repeat collimator centering scan? --may be we can skip this
High current checkout
- Establish high current to hall dump. 150uA. Radiation, IC levels should match expectation.
- Ion Chamber calibration - thick Target 0.25 shift
- Should be easy/standard procedure, but verify calibration matches old calibration
- verify trip points match old trip points, running conditions for high current match old high current running. If not, differences must be understood.
- Setup beam through Compton chicane.
- must keep watch on Compton rates whenever through Compton chicane
- FFB commissioning. Expected to be used when running high current
- Establish beam on cold target
- integrating mode detectors
- C-hole, spot++
- Thick carbon target. Spot++ to verify on target
- Thick carbon to high current. Burn off (30 minutes each at 50,100,150uA)
- Ca-48 target. Spot++ to verify on target
- Ca-48 to high current. Burn off (30 minutes each at 50,100,150uA)
- Verify beam modulation functionality
- low current test, then high current to watch of Ion Chamber rates or beam loss
- Ignore orthogonality for first test
- Test production with 1-2 hours.
- detector and SAM widths
- regression sensitivities
- beam noise and asymmetry check
- collimator water and thermocouple temperatures,
- radiation monitors and ion chamber stability
- Tune beam modulation orthogonality, if needed, after 1-2 hours, then take at least 1 hour data to confirm test.
Start Production
Other Early Tests and Checks (first few days)
- Beam Monitor calibration
- soon after restart
- Run current scan using Unser to calibrate beam charge monitor readout in parity DAQ.
- Run the Compton DAQ concurrently, for calibration
- Go to FixedGain when at 150 uA on bpms (list of crates???) and repeat local current scan to set bpm pedestals.
- Fine-tune detector alignment
- At detector alignment
- Compton commissioning
- Moller measurement
Other Tests and Checks
- Do we need to repeat spin-dance?
- Linearity Studies
- Repeat Q2 and pointing measurements
- Arc Energy Measurement
- Background Studies
- Scans of Septum Magnet and HRS Dipole
- Possible scan of Q1 to optimize acceptance and verify that collimator defines acceptance. (old idea, still relevant?)
Polarimetry Studies
- Compton Commissioning
- Beam Tune, Background reduction
- Compton Cavity Checkout
- Establish laser in cavity
- Move photon detector to visual center, scan beam for collisions
- Establish orbit lock
- Photon Detector Checkout
- Detector power-up, find optimal gain.
- Center detector with fingers.
- 15 minute spectrum and rate study.
- Brief target-out run? (Probably not necessary but may be desirable)
- Polarization commissioning run (up to 2 hour target-in)
- Jaws checkout (?)
- Moller Commissioning -- Simona Malace, Don Jones, et.al.