Difference between revisions of "DAQ Testing/20190118"

From PREX Wiki
Jump to navigationJump to search
Line 1: Line 1:
 
Using batteries to look for pedestal noise floor, drift, or helicity correlated pickup
 
Using batteries to look for pedestal noise floor, drift, or helicity correlated pickup
 
* Run 1032 - failed
 
* Run 1032 - failed
* Run 1033 - Battery (1.6 V) plugged into RHRS first ADC, channel 0 - 5300 events at 30 Hz flip rate, 500 us Tsettle time (scalers and synch monitor are now invalid - were set for 240 Hz and 160 us Tsettle timing)
+
* Run 1033 - Battery (1.6 V) plugged into RHRS first ADC, channel 7 (index 7, 8th channel) - 5300 events at 30 Hz flip rate, 500 us Tsettle time (scalers and synch monitor are now invalid - were set for 240 Hz and 160 us Tsettle timing)
* Run 1034 - Battery (1.6 V) plugged into RHRS second ADC, channel 3 - 5300 events (I may have accidentally moved lumi signal in channel 6 to 5, either way, its in 5 now)
+
* Run 1034 - Battery (1.6 V) plugged into RHRS second ADC, channel 4 - 5300 events (I may have accidentally moved lumi signal in channel 1 to 2, either way, its in 2 now)
* Run 1035 - Battery (1.6 V) plugged into LHRS only ADC, channel
+
* Run 1035 - Battery (1.6 V) plugged into LHRS only ADC, channel 7 -

Revision as of 16:19, 18 January 2019

Using batteries to look for pedestal noise floor, drift, or helicity correlated pickup

  • Run 1032 - failed
  • Run 1033 - Battery (1.6 V) plugged into RHRS first ADC, channel 7 (index 7, 8th channel) - 5300 events at 30 Hz flip rate, 500 us Tsettle time (scalers and synch monitor are now invalid - were set for 240 Hz and 160 us Tsettle timing)
  • Run 1034 - Battery (1.6 V) plugged into RHRS second ADC, channel 4 - 5300 events (I may have accidentally moved lumi signal in channel 1 to 2, either way, its in 2 now)
  • Run 1035 - Battery (1.6 V) plugged into LHRS only ADC, channel 7 -