Back to [[Main_Page|Main Page]] >> [[DAQ_Doc_Portal|DAQ Documentation Portal]] >> [[DAQ_Testing|DAQ Testing]] >> [[DAQ_Commissioning_Notes|DAQ Commissioning Notes]] | Back to [[Main_Page|Main Page]] >> [[DAQ_Doc_Portal|DAQ Documentation Portal]] >> [[DAQ_Testing|DAQ Testing]] >> [[DAQ_Commissioning_Notes|DAQ Commissioning Notes]] | ||
− | [[DAQ_Testing/20190502|Previous Day of Testing]] << >> [[DAQ_Testing/ | + | [[DAQ_Testing/20190502|Previous Day of Testing]] << >> [[DAQ_Testing/20190508|Next Day of Testing]] |
May 4th, 2019 | May 4th, 2019 |
Latest revision as of 18:41, 8 May 2019
Back to Main Page >> DAQ Documentation Portal >> DAQ Testing >> DAQ Commissioning Notes
Previous Day of Testing << >> Next Day of Testing
May 4th, 2019 Testers: Cameron Clarke
Status
- Test new 7200 scaler
- Add HAPTB to INJ
New 7200
- I finished testing the 7200 in the injector (straight swap) - it works - run 2153
- I set the original one back in after - still works (run 2154)
INJ HAPTB
- I added a HAPTB to INJ crate - it hasn't been configured yet, should match CH when done
- I added 16 bit DAC connection into vqwk_10_05 channel
- HAPTB Tsettle LNE required splitting the SIS3801's VETO into a 726 trans - looks good on scope...