== Counting to Integrating == | == Counting to Integrating == | ||
− | # Setting HV | + | # Setting HV - this is important |
#* VDC HV off. This is important. | #* VDC HV off. This is important. | ||
#* Scintillator HV off. This is important. | #* Scintillator HV off. This is important. | ||
− | #* Info about detector HV | + | #* [http://hallaweb.jlab.org/parity/prex/runinfo/opmanual.html#DetectorHV Info about detector HV] |
+ | |||
#* Adjust Quartz Detector HV for Integrating mode. | #* Adjust Quartz Detector HV for Integrating mode. | ||
# Switching to High-Current Integrating Mode | # Switching to High-Current Integrating Mode | ||
Line 21: | Line 22: | ||
#* Septum on | #* Septum on | ||
# Set Up Beam | # Set Up Beam | ||
− | #* Do the beam recovery procedure | + | #* Do the [http://hallaweb.jlab.org/parity/prex/runinfo/How_to_recover_beam.html beam recovery procedure] |
# Take data ! Check all data including cavities. | # Take data ! Check all data including cavities. | ||
Revision as of 11:10, 17 June 2019
In the process of editing and updating this page. Please contact Juliette Mammei crowder@jlab.org with questions or suggestions.
PREX Main<< HOW TOs
This page details the checklists for switching between counting and integrating modes. For reference, here are the instructions from PREX I
Counting to Integrating
- Setting HV - this is important
- VDC HV off. This is important.
- Scintillator HV off. This is important.
- Info about detector HV
- Adjust Quartz Detector HV for Integrating mode.
- Switching to High-Current Integrating Mode
- Septum off
- Take access.
- Expert to do the switch of the detector signals to integrating mode
- Cavities: take out amplifiers. Put in attenuator on BCM. This is important.
- Set up cavities pre-gain and phase using established values.
- Septum on
- Set Up Beam
- Do the beam recovery procedure
- Take data ! Check all data including cavities.
Integrating to Counting
- Preliminaries. to do long before switching
- Make sure spectrometer DAQ works (might need reboots). Use the CODA guide if necessary.
- Check cavities in integration mode and counting mode at high current. Run analysis macros. Do they work ?
- Check with spot++ that rastered size is approx 4 x 3 mm (whatever is our present norm.)
- Check the beam position ("centered" to within 0.3 mm, using the production criterea, see white board).
- Switching to Low-Current Counting Mode
- Septum off
- Take access.
- Expert to do the switch of the detector signals to counting mode
- Switch cavities to have amplifiers. Take out attenuator on BCM. Note: Cannot exceed 2 uA with amplifiers in place.
- Set up cavities pre-gain and phase using established values.
- Septum on
- Set Up Beam
- Do the beam recovery procedure.
- Establish 50 nA on Pb target (or whatever current on another target). Normally with laser attenuator = 65 and slit = 13, and then some adjustment to get rates low.
- Setting HV
- Do NOT turn on scint or VDC HV yet ! Must establish rates first.
- Info about detector HV
- Turn on the HV for upper scint and S0 (if S0 is in place).
- Adjust Quartz Detector HV for Counting mode.
- VDC HV on (4 kV) if the T1/T5 rates are below 500 kHz. If above 500 kHz do not turn on HV.
- Take CODA run and analyze it as usual.