== Beamline instrumentation calibration == | == Beamline instrumentation calibration == | ||
#Check the following has been performed | #Check the following has been performed | ||
− | + | #* 50uA-70uA beam current, through Compton (backgrounds on fingers and Compton photon detector) with no target | |
− | #* 50uA-70uA beam current, through Compton (backgrounds on fingers and Compton photon detector) | ||
#* Establish 4x4 raster. Insert lead target. | #* Establish 4x4 raster. Insert lead target. | ||
− | #* Detector and SAM voltages (Dustin M) | + | #* Set Detector and SAM voltages (Dustin M) |
+ | |||
+ | #Check Conditions OK | ||
+ | #*collimator water | ||
+ | #*thermocouple temperatures | ||
+ | #*radiation monitors | ||
+ | #*ion chamber stability | ||
+ | |||
+ | #Setup Locks (these are necessary for main detectors and SAM calibration) | ||
+ | #*Slow Target Lock on 4A, 4E (jmenu->"slow locks"-> Orbit Locks Expanded View -> HallATGT) | ||
+ | #* Fast feedback for HallA on (iocse9) | ||
+ | #* If beam is going through Compton, HallAComLock should probably be on | ||
+ | #* Should HallAMolLock be on? | ||
+ | |||
+ | #Coordinate | ||
+ | #* Should Compton and Moller DAQ be running during calibration? | ||
+ | #* Dustin and Devi examining SAMs and Detectors with online analyzer during calibration | ||
+ | #* Juan Carlos/Amali/Adam monitor Compton background rates and run the DAQ? | ||
+ | |||
+ | #Calibration Procedure A: with UNSER | ||
+ | #*start at 70uA for 1min (scandata1==value in BCM0l02) | ||
+ | #*beam off | ||
+ | #*wait for 5sec(for unser to equilibrate), take unser pedestal data point for 30sec (scandata1=0) | ||
+ | #*beam on at 75uA | ||
+ | #*check slow target lock is in position | ||
+ | #*check FFB on for HallA | ||
+ | #*check no bpms are saturated on iocse9 | ||
+ | #*wait for 5sec(for unser to equilibrate), take next data point for 1min | ||
+ | #*repeat for 65uA,60uA,55uA,50uA,40uA,30uA,20uA,10uA | ||
+ | |||
+ | #Calibration Procedure B: with pre-calibrated Hall BCM (analog upstream bcm) | ||
+ | #*start at 70uA for 1min (scandata1==value in BCM0l02) | ||
+ | #change current to 75uA | ||
+ | #*check slow target lock is in position | ||
+ | #*check FFB on for HallA | ||
+ | #*check no bpms are saturated on iocse9 | ||
+ | #*take next 1min data point | ||
+ | #*repeat for 65uA,60uA,55uA,50uA,40uA,30uA,20uA,10uA | ||
+ | |||
+ | #Calibration Procedure C: for BPMs (Compton bpms not yet calibrated) | ||
+ | #*at 70uA, set gains off for iocse9,iocse10 (Hall) iocse11,iocse12,iocse19,iocinse4 (Inj) | ||
+ | #*rperform Calibration Procedure B with pre-calibrated Hall BCM | ||
# <font color="blue">Beamline instrumentation calibration</font> <span style="background-color:#F1C40F">0.5 shift</span> Caryn P, | # <font color="blue">Beamline instrumentation calibration</font> <span style="background-color:#F1C40F">0.5 shift</span> Caryn P, |
Revision as of 09:25, 25 June 2019
Beamline instrumentation calibration
- Check the following has been performed
- 50uA-70uA beam current, through Compton (backgrounds on fingers and Compton photon detector) with no target
- Establish 4x4 raster. Insert lead target.
- Set Detector and SAM voltages (Dustin M)
- Check Conditions OK
- collimator water
- thermocouple temperatures
- radiation monitors
- ion chamber stability
- Setup Locks (these are necessary for main detectors and SAM calibration)
- Slow Target Lock on 4A, 4E (jmenu->"slow locks"-> Orbit Locks Expanded View -> HallATGT)
- Fast feedback for HallA on (iocse9)
- If beam is going through Compton, HallAComLock should probably be on
- Should HallAMolLock be on?
- Coordinate
- Should Compton and Moller DAQ be running during calibration?
- Dustin and Devi examining SAMs and Detectors with online analyzer during calibration
- Juan Carlos/Amali/Adam monitor Compton background rates and run the DAQ?
- Calibration Procedure A: with UNSER
- start at 70uA for 1min (scandata1==value in BCM0l02)
- beam off
- wait for 5sec(for unser to equilibrate), take unser pedestal data point for 30sec (scandata1=0)
- beam on at 75uA
- check slow target lock is in position
- check FFB on for HallA
- check no bpms are saturated on iocse9
- wait for 5sec(for unser to equilibrate), take next data point for 1min
- repeat for 65uA,60uA,55uA,50uA,40uA,30uA,20uA,10uA
- Calibration Procedure B: with pre-calibrated Hall BCM (analog upstream bcm)
- start at 70uA for 1min (scandata1==value in BCM0l02)
- change current to 75uA
- check slow target lock is in position
- check FFB on for HallA
- check no bpms are saturated on iocse9
- take next 1min data point
- repeat for 65uA,60uA,55uA,50uA,40uA,30uA,20uA,10uA
- Calibration Procedure C: for BPMs (Compton bpms not yet calibrated)
- at 70uA, set gains off for iocse9,iocse10 (Hall) iocse11,iocse12,iocse19,iocinse4 (Inj)
- rperform Calibration Procedure B with pre-calibrated Hall BCM
- Beamline instrumentation calibration 0.5 shift Caryn P,
- Current ramp for BPM/BCM calibrations and detector pedestals detailed plan
- Initial production condition test 0.25 shift Caryn P,
- detector and SAM widths
- regression sensitivities
- beam noise and asymmetry check
- collimator water and thermocouple temperatures,
- radiation monitors and ion chamber stability
- Width studies 0.5 shift KK,
- detector and SAM widths, targets, shutters, etc
- deinstall shutters at conclusion of test