Detailed Calibration plan
From PREX Wiki
Revision as of 11:00, 25 June 2019 by Caryn Palatchi (talk | contribs)
Beamline instrumentation calibration
- Check the following has been performed
- 50uA-70uA beam current, through Compton (backgrounds on fingers and Compton photon detector) with no target
- Establish 4x4 raster. Insert lead target. (low current, ramp up. check collimator water, vary raster size 4x6?)
- Set Detector and SAM voltages (Dustin M)
- Check Conditions OK
- collimator water
- thermocouple temperatures
- radiation monitors
- ion chamber stability
- Setup Locks (these are necessary for main detectors and SAM calibration)
- Slow Target Lock on 4A, 4E (jmenu->"slow locks"-> Orbit Locks Expanded View -> HallATGT)
- Fast feedback for HallA on (iocse9)
- If beam is going through Compton, HallAComLock should probably be on
- Should HallAMolLock be on?
- Coordinate
- Should Compton and Moller DAQ be running during calibration?
- Dustin and Devi examining SAMs and Detectors with online analyzer during calibration
- Juan Carlos/Amali/Adam monitor Compton background rates and run the DAQ?
- Ye puts Cavity BPMs in high current mode (opsl01 -> menubar->EDM->BPM->BPM Diagnostics DR(12GeV)->HallA BCMs B,C,D -> button -> Expert Controls -> HWReg button -> DACSL1 = 0x1 / 0x1 and DACSL2 = 0x1 / 0x1 -> back out to previous screen XVGA2 = 0 and YVGA2 =0 (for low current do 0x0 and VGA2=30). Do for all 3 BPM Cavs.
- Calibration Procedure A: with UNSER
- start at 70uA for 1min (scandata1==value in BCM0l02)
- beam off
- wait for 5sec(for unser to equilibrate), take unser pedestal data point for 30sec (scandata1=0)
- beam on at 75uA
- check slow target lock is in position
- check FFB on for HallA
- check no bpms are saturated on iocse9
- wait for 5sec(for unser to equilibrate), take next data point for 1min
- repeat for 65uA,60uA,55uA,50uA,40uA,30uA,20uA,10uA
- Calibration Procedure B: with pre-calibrated Hall BCM (analog upstream bcm)
- start at 70uA for 1min (scandata1==value in BCM0l02)
- change current to 75uA
- check slow target lock is in position
- check FFB on for HallA
- check no bpms are saturated on iocse9
- take next 1min data point
- repeat for 65uA,60uA,55uA,50uA,40uA,30uA,20uA,10uA
- Calibration Procedure C: for BPMs (Compton bpms not yet calibrated)
- at 70uA, set gains off for iocse9,iocse10 (Hall) iocse11,iocse12,iocse19,iocinse4 (Inj)
- perform Calibration Procedure B with pre-calibrated Hall BCM