August 4th, 2020 Tester: Cameron Clarke

Replacing VQWK number 025 from L:HRS _0_3 bad channel

To make these plots I have used the DAQ_Checkout.conf decoding configuration (for the CH_INJ CODA config) and used the panguin macro QualityAssurance/testChannels.cfg

Checking out all ADC modules

Using the RampAll config (rampdac16 setting in CH crate cedit config)

  • Run 7537 - Looks good by eye, if interested see PDFs in ~/cameronc/DAQ-calibration/CREX-August-Restart/
  • Run 7538-7544 : remaining 7 ADCs in CH crate, all appear good
  • Run 7545 - added rampdac16 to L and RHRS in CEDIT - test run, works