August 17 2020
Daily Meeting BlueJeans
Production Run List - Slug 192 - IHWP IN - Flip-RIGHT
Old Alignment
- Good - Run 7776 - Small 12X bump once, add it to cut out > 0mm 12X cut with run 7778. Prior alignment configuration, meaning it will need a bmod slopes sub-segment. Looks good.
- Good - Run 7778 - looks good. Add BPM 12X cut out > 0.0mm. Uses prior alignment setting, meaning it needs a beamMod subsegment
New detectors alignment
- Good - Run 7779 - looks good. First run with the new alignment - perfect match of L and R HRS positioning. Extend BPM 12X cut out > 0.0 mm from prior runs.
- Good - Run 7780 - looks good. Second run with the new alignment - perfect match of L and R HRS positioning. Extend BPM 12X cut out > 0.0 mm from prior runs.
- Good - Run 7781 - looks good. Extend BPM 12X cut out > 0.0 mm from prior runs.
- Good - Run 7782 - looks good. Extend BPM 12X cut out > 0.0 mm from prior runs.
- Good - Run 7783 - looks good
- Good - Run 7784 - looks good
- Good - Run 7785 - looks good. Needs 12X cuts. Robert is looking into adding a generic burp cut
Came back to 192
- Good - Run 7799 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7800 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7801 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7802 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7803 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7804 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7805 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
- Good - Run 7806 - Looks good. Less 12X motion. Back from long down
Production Run List - Slug 193 - IHWP OUT - Flip-RIGHT
- Good - Run 7786 - looks good. Cut out 12X BPM > 0.0mm
- Good - Run 7787 - looks good.
- Good - Run 7788 - looks good. Add BPM 12X cuts to remove > 0.0 mm and < -1.5 mm
- Good - Run 7789 - looks good. There is noisy hair in BPM Ys. There is a big spike in BPM 12X, so cut out > 0.0mm and < -1.5 mm
- Good - Run 7790 - looks good. There is noisy hair in BPM Ys. There is a big spike in BPM 12X, so cut out > 0.0mm and < -1.5 mm
- Good - Run 7791 - looks good
- Good - Run 7792 - Usr yield and Asym have spikey noise, coming from BPM 12X noise. Cut out BPM 12X > -0.25
- Good - Run 7793 - looks good. Some hairy BPM 12X, but small
- Good - Run 7794 - looks good. Some hairy BPM 12X, but small
- Good - Run 7795 - looks good. Some hairy BPM 12X, but small
- Good - Run 7796 - looks good. Some hairy BPM 12X, but small
- Good - Run 7797 - looks good. BPM12X noise - cut out >-0.25, < -1.5 mm
Non-Production Runs and Bad Runs
Issues
- Running with FFB off
- Victoria noticed that coil 6 in BPM Ys was getting caught by the burp cut. I've retroactively removed the BPM Y burp cuts from the open run ranged files (which is annoying and possibly a problem... but it was already being removed for each run that had a 12X cut, which so far is like half of them)
Updates
- BMOD tune seems OK.