Flipper

From PREX Wiki
Revision as of 21:35, 1 August 2019 by Caryn Palatchi (talk | contribs)
Jump to navigationJump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
  • Flip IHWP state
    • to flip IHWP state, will have to inform other halls of intention to flip state first. Then call MCC and ask them to flip state.
    • Note: After the IHWP state has been changed, the parity alarms will go off until the voltages are set correctly.
    • Then (after IHWP is in new desired state), on apar@adaq2, in ~/feedback2019, run ./FlipPCValues to set the voltages correctly
    • Double check voltages are correct: on apar@adaq2, in /feedback2019, run ./checkPCvalues The output should tell you the voltages are good, otherwise call an expert.
    • if alarms go off and values don't match those in table, call an Expert
    • Update the slug number and IHWP state of the white board, write down the time of flip and first run number for new slug.
    • Check the values against the table in Parity Checks

Special note for slugs 24,25

  • For slug 24 and 25 on 8/1/2019, use ./CancelPCValues and ./checkCancelPCValues
    • to flip IHWP state, will have to inform other halls of intention to flip state first. Then call MCC and ask them to flip state.
    • Note: After the IHWP state has been changed, the parity alarms will go off until the voltages are set correctly.
    • Then (after IHWP is in new desired state), on apar@adaq2, in ~/feedback2019/CancelFlip, run ./CancelPCValues to set the voltages correctly
    • Double check voltages are correct: on apar@adaq2, in /feedback2019/CancelFlip, run ./checkCancelPCvalues The output should tell you the voltages are good, otherwise call an expert.
    • Update the slug number and IHWP state of the white board, write down the time of flip and first run number for new slug.
    • Double check AposV which has been modified: Parity Alarm for AposV will still be going off. It is OK to silence it ONLY for these slugs as long as for IHWPin, AposV=15 and for IWHPout, AposV=-237.