Line 24: | Line 24: | ||
* Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3595_000 run3595] | * Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3595_000 run3595] | ||
* Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3596_000 run3596] | * Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3596_000 run3596] | ||
− | * | + | * Bad - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3598_000 run3598] |
− | + | ||
− | + | ||
− | + | ||
− | + | ||
− | + | ||
** rcdb says current scan but BCM yields are ~70uA, RMS on detectors look OK! | ** rcdb says current scan but BCM yields are ~70uA, RMS on detectors look OK! | ||
** 128 ppm charge asymmetry in BCMs, no Aq running ? -> There are 3 periods of different avg Aq, so maybe a cut will cut out the very large bad Aq stuff | ** 128 ppm charge asymmetry in BCMs, no Aq running ? -> There are 3 periods of different avg Aq, so maybe a cut will cut out the very large bad Aq stuff | ||
** BPM 4ac still saturating and 4eX still 12 um widths | ** BPM 4ac still saturating and 4eX still 12 um widths | ||
− | * | + | * Bad - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3601_000 run3601] |
** Mini-run 0 has ~4X RMS (det reg_asym). Note: Aq feedback off, RHWP rotated back to nominal value | ** Mini-run 0 has ~4X RMS (det reg_asym). Note: Aq feedback off, RHWP rotated back to nominal value | ||
** There is large BPM noise (20 um widths in the online plots, though it may be some strange outliers) | ** There is large BPM noise (20 um widths in the online plots, though it may be some strange outliers) | ||
Line 42: | Line 37: | ||
* Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3606_000 run3606] | * Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3606_000 run3606] | ||
* Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3607_000 run3607] | * Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3607_000 run3607] | ||
− | * | + | * Good - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3608_000 run3608] |
** Not sure why the run was too short | ** Not sure why the run was too short | ||
+ | ** Too big Aq | ||
== Non-Production Runs and Bad Runs == | == Non-Production Runs and Bad Runs == | ||
Line 56: | Line 52: | ||
== Pedestal/Calibrations Updates == | == Pedestal/Calibrations Updates == | ||
* DET HV Calibration [https://hallaweb.jlab.org/parity/prex/onlinePlots/summary/2019-07-24/run3594_000/run3594_000_summary_main_detector.pdf run3594] | * DET HV Calibration [https://hallaweb.jlab.org/parity/prex/onlinePlots/summary/2019-07-24/run3594_000/run3594_000_summary_main_detector.pdf run3594] | ||
+ | * Calibration - [https://hallaweb.jlab.org/parity/prex/onlinePlots/index.php?run=summary/all/run3597_000 run3597] | ||
+ | ** RMS (after regression) is fluctuating between miniruns. Looks like current scan run! | ||
+ | ** It is a current scan, bad Aq | ||
+ | ** The BPM4ac is saturating the whole time | ||
+ | ** The BPM 4eX is very wide (12 um widths) | ||
* PITA Scan RHWP=500 [https://hallaweb.jlab.org/parity/prex/onlinePlots/summary/2019-07-25/run3599_000/run3599_000_all.pdf run3599] | * PITA Scan RHWP=500 [https://hallaweb.jlab.org/parity/prex/onlinePlots/summary/2019-07-25/run3599_000/run3599_000_all.pdf run3599] | ||
* PITA Scan RHWP=500 [https://hallaweb.jlab.org/parity/prex/onlinePlots/summary/2019-07-25/run3600_000/run3600_000_all.pdf run3600] | * PITA Scan RHWP=500 [https://hallaweb.jlab.org/parity/prex/onlinePlots/summary/2019-07-25/run3600_000/run3600_000_all.pdf run3600] | ||
+ | ** Aq feedback is off, very short run with elevated det width, big Aq mean | ||
== Issues == | == Issues == |
Latest revision as of 14:42, 25 September 2019
Documentation |
HOW TOs for shift crew |
Expert Tools |
All Expert Contacts |
PREX Main << Weekly Analysis Coordinator << WAC Notes << Instructions
WAC Notes July 23 2019 << >> WAC Notes July 25 2019
Contents
July 24 2019
Production Run List
Start Slug 12, IHWP In, FFB on always
- Good - run3595
- Good - run3596
- Bad - run3598
- rcdb says current scan but BCM yields are ~70uA, RMS on detectors look OK!
- 128 ppm charge asymmetry in BCMs, no Aq running ? -> There are 3 periods of different avg Aq, so maybe a cut will cut out the very large bad Aq stuff
- BPM 4ac still saturating and 4eX still 12 um widths
- Bad - run3601
- Mini-run 0 has ~4X RMS (det reg_asym). Note: Aq feedback off, RHWP rotated back to nominal value
- There is large BPM noise (20 um widths in the online plots, though it may be some strange outliers)
- Good - run3602
- Good - run3603
- Good - run3604
- Good - run3605
- Good - run3606
- Good - run3607
- Good - run3608
- Not sure why the run was too short
- Too big Aq
Non-Production Runs and Bad Runs
- Junk run3589
- Junk run3590
- Junk run3591
- Test 50uA run3592
- Test 50uA run3593
- Test run @50uA large Aq mean
- Junk run3609
Pedestal/Calibrations Updates
- DET HV Calibration run3594
- Calibration - run3597
- RMS (after regression) is fluctuating between miniruns. Looks like current scan run!
- It is a current scan, bad Aq
- The BPM4ac is saturating the whole time
- The BPM 4eX is very wide (12 um widths)
- PITA Scan RHWP=500 run3599
- PITA Scan RHWP=500 run3600
- Aq feedback is off, very short run with elevated det width, big Aq mean
Issues
- Starting from the beginning of production running analysis fails all events without any indication why HALOG
- Analog BCMs are too noisey for recent runs HALOG
- Regression seems to be failing for run3604 run3605HALOG
- Double peaks and shoulders in bcm dd run3604 run3605 run3603 HALOG
- Residual correlation with AqDs3 run3605 run3604 run3603 run3602 run3601
- Seems all the runs have fluctuation on RMS on combined asymmetry (postpan).
Other Status Updates
- Accumulated charge = 12.65C
- Changed regression set (and added combined BPM 4a_4ec X) - git commit - HALOG
- new file: prex_combiner.3583-.map
- new file: prex_corrolator.3583-.conf
- Updated postpan regression "combo_reg.3583-.conf" - git commit 3f8fa67 (prompt repo commit aea4bb3), HALOG
- Added aggregator scripts to prompt repo - commit 6838abbbae9ca3e02a3a2d6e8d29050b5f9d2437
- Updated priliminary pedestal maps for 70uA running (for Main, AT, and SAM detectors) HALOG