Difference between revisions of "Beam and Hall checks"
From PREX Wiki
Jump to navigationJump to searchLine 23: | Line 23: | ||
** | ** | ||
* Raster size (x:y mm, see log 3654466 )? | * Raster size (x:y mm, see log 3654466 )? | ||
+ | ** Go to JMenu, search "Hall A Fast Raster", Click on Screens->Hall A Fast Raster | ||
+ | ** If the raster is on then the "Set Raster Size" section will have height and width readback values non-zero | ||
+ | ** Also we should check the X-Y raster scope in the center DAQ room | ||
* Fast feedback on? Yes/No | * Fast feedback on? Yes/No | ||
* Beam Position Monitor X/Y (mm) A | * Beam Position Monitor X/Y (mm) A |
Revision as of 17:19, 12 June 2019
In the process of editing and updating this page. Please contact Juliette Mammei crowder@jlab.org with questions or suggestions.
PREX Main<< Information for Shift Takers << Shift Leader << Shift Checklist
Alarm Handler
- Is alarm handler running on hacweb7?
- On hacweb 7, you should see a window in the upper right corner of the screen
- Look first in the "Alarm Handler" desktop to see if it is there
- Check all the windows if not
- If not open, see Alarm Handler
- On hacweb 7, you should see a window in the upper right corner of the screen
- Is the volume up on hacweb7? (both hardware and software)
- Beam Energy (GeV)
- Beam current (µA)
- Downstream raster (B) on? Yes/No (???)
- picture of scope when raster is off, and when raster is on
- Raster size (x:y mm, see log 3654466 )?
- Go to JMenu, search "Hall A Fast Raster", Click on Screens->Hall A Fast Raster
- If the raster is on then the "Set Raster Size" section will have height and width readback values non-zero
- Also we should check the X-Y raster scope in the center DAQ room
- Fast feedback on? Yes/No
- Beam Position Monitor X/Y (mm) A
- Beam Position Monitor X/Y (mm) E (BPM B in Hall A Tools)
- BCM temperature (K) (PV, PS, Thermocouple) If these differ by +/-.5 from (110.0,110.0,115.0) (Fahrenheit) contact RC and halog
- These temperature readbacks are located in the middle DAQ room, in rack CH0B06 (far left) at the very top
- It is important that these temperatures stay stable within 0.5 Fahrenheit or else the BCM current readout will become uncalibrated
- Arc energy NMR locked?
- Go to JMenu, search "Energy Lock", click on applications->Energy Lock
- If the energy is locked all 3 lights should be green
- VESDA System OK? (VESDA is the smoke/fire detection and alarm system. Check the Hall A VESDA Console on the wall behind the door
- Current Target
- Argon pressure (psi)
- Ethane pressure (psi)
- CO2 pressure (psi)
- Visual Hall Inspection OK?
- Septum current (A) and polarity?
- Upstream/downstream corrector currents (A) ?
- Check the scalers