Difference between revisions of "Beam and Hall checks"
From PREX Wiki
Jump to navigationJump to searchLine 7: | Line 7: | ||
[[File:Raster_bad.jpg|thumb|600px|right|Figure 3 - A picture of the scope when the raster is off. This would be BAD if beam were on!]] | [[File:Raster_bad.jpg|thumb|600px|right|Figure 3 - A picture of the scope when the raster is off. This would be BAD if beam were on!]] | ||
[[File:BCM_temps.jpg|thumb|600px|right|Figure 4 - Picture of the BCM temperature readout in the computer room.]] | [[File:BCM_temps.jpg|thumb|600px|right|Figure 4 - Picture of the BCM temperature readout in the computer room.]] | ||
− | [[File:Halltools_gas.png|thumb|600px|right|Figure | + | [[File:Fast_feedback_off.png|thumb|600px|right|Figure 5 - Snapshot of the Fast Feedback Controls when the FFB is <b>OFF</b>.]] |
− | [[File:Halltools_other.png|thumb|600px|right|Figure | + | [[File:Halltools_gas.png|thumb|600px|right|Figure 6 - The gas section in the Hall A Tools window.]] |
− | [[File:VESDA.jpg|thumb|600px|right|Figure | + | [[File:Halltools_other.png|thumb|600px|right|Figure 7 - Camera and alignment info in the Hall A Tools window.]] |
− | [[File:Camera_location.jpg|thumb|600px|right|Figure | + | [[File:VESDA.jpg|thumb|600px|right|Figure 8 - The VESDA panel in the counting house.]] |
− | [[File:Camera_controls.jpg|thumb|600px|right|Figure | + | [[File:Camera_location.jpg|thumb|600px|right|Figure 9 - Where you can find the camera display and the controls in the counting house.]] |
+ | [[File:Camera_controls.jpg|thumb|600px|right|Figure 10 - The camera controls.]] | ||
Line 57: | Line 58: | ||
=== Gasses === | === Gasses === | ||
− | * Pressures (see Figure | + | * Pressures (see Figure 6) |
** Argon pressure (psi) | ** Argon pressure (psi) | ||
** Ethane pressure (psi) | ** Ethane pressure (psi) | ||
** CO2 pressure (psi) | ** CO2 pressure (psi) | ||
− | * VESDA System OK? (see Figure | + | * VESDA System OK? (see Figure 8) |
** VESDA is the smoke/fire detection and alarm system | ** VESDA is the smoke/fire detection and alarm system | ||
** Check the Hall A VESDA Console on the wall behind the door | ** Check the Hall A VESDA Console on the wall behind the door | ||
Line 69: | Line 70: | ||
=== Hall/Counting House === | === Hall/Counting House === | ||
− | * Visual Hall Inspection OK? (see Figure | + | * Visual Hall Inspection OK? (see Figure 9) |
− | ** Make sure to move the camera around and zoom in (See Figure | + | ** Make sure to move the camera around and zoom in (See Figure 10) |
** when complete, zoom all the way back out | ** when complete, zoom all the way back out | ||
* Current Target (snapshot) | * Current Target (snapshot) |
Revision as of 10:41, 18 June 2019
In the process of editing and updating this page. Please contact Juliette Mammei crowder@jlab.org with questions or suggestions.
PREX Main<< Information for Shift Takers << Shift Leader << Shift Checklist
Alarm Handler
- Is alarm handler running on hacweb7?
- On hacweb 7, you should see a window in the upper right corner of the screen
- Look first in the "Alarm Handler" desktop to see if it is there
- Check all the windows if not
- If not open, see Alarm Handler
- On hacweb 7, you should see a window in the upper right corner of the screen
- Is the volume up on hacweb7? (both hardware and software)
Beamline
- Beam Energy (GeV) (see Figure 2)
- Beam Current (µA)
- Beam Position Monitor (BPM) X/Y (mm) A
- Beam Position Monitor (BPM) X/Y (mm) E (Called BPM B in Hall A Tools)
- Downstream raster (B) on? Yes/No (upstream raster is not in use)
- when raster is off, looks like Figure 3
- when raster is on, something like halog 3654466
- BCM temperature (K) (PV, PS, Thermocouple)
- If these differ by +/-.5 from (110.0, 110.0, 115.0) (Fahrenheit) contact RC and halog
- These temperature readbacks are located in the middle DAQ room, in rack CH0B06 (far left) at the very top (see Figure 4)
- It is important that these temperatures stay stable within 0.5 Fahrenheit or else the BCM current readout will become uncalibrated
- Raster size (x:y mm, see log 3654466)?
- Fast feedback (FFB) on? Yes/No
- To check whether the FFB is on or off
- in a terminal window, type JMenu
- search for "fast"
- Choose "Fast Feedback Control" (see Figure , which is for FFB OFF)
- The status is indicated by the button that is depressed (OFF and RUN in the figure)
- FFB should be ON during normal production running!
- However, FFB will be paused during beam modulation
- To check whether the FFB is on or off
- Arc energy NMR locked?
- Go to ARC or the "Field Management Screen"
- Verify if the box locked is ON/OFF
Gasses
- Pressures (see Figure 6)
- Argon pressure (psi)
- Ethane pressure (psi)
- CO2 pressure (psi)
- VESDA System OK? (see Figure 8)
- VESDA is the smoke/fire detection and alarm system
- Check the Hall A VESDA Console on the wall behind the door
Hall/Counting House
- Visual Hall Inspection OK? (see Figure 9)
- Make sure to move the camera around and zoom in (See Figure 10)
- when complete, zoom all the way back out
- Current Target (snapshot)
- Septum current (A) and polarity?
- Upstream/downstream corrector currents (A) ?
Next in checklist >> Parity Checks