Difference between revisions of "TEDf-VQWK-Testing"

From PREX Wiki
Jump to navigationJump to search
Line 69: Line 69:
 
* ADC 20
 
* ADC 20
 
** Determined not to work - 2/13/2019
 
** Determined not to work - 2/13/2019
**  
+
** Channels 4-7 (top 4) do not work
 +
** Signal ramps to -10 V for all events, regardless of containing a voltage signal or not
 
* ADC 26
 
* ADC 26
 
** Determined to work - 2/14/2019
 
** Determined to work - 2/14/2019
 +
* ADC 31
 +
** Determined not to work - 2/15/2019
 +
** There are ROC read errors for 1/3 events or so
 +
** All channels have randomly stepping noisy signal that rails into certain regions and no clear response to input voltage
 +
** ROC Read errors get worse when plugging a signal into a channel
 +
** Some channels show differing response to input... characterizing problems will be a bit hard, so assume normal input stage problems? Or wiring to the later stages is bad across all channels?
 +
* ADC 32
 +
** Determined not to work - 2/14/2019
 +
** All 8 channels have interesting problems
 +
** It looks like regular bad amplifier in input stage problems
 +
** But, in the third run I took with this board it looks like the signal stopped being read by digital components entirely (run 1662, all signals == 0)
 +
* ADC 33
 +
** Determined not to work - 2/15/2019
 +
** Regular bad channels
 +
** Channel 6 "works" but probably it would be best to go ahead and give it a standard replacement treatment too?
 +
* ADC 37
 +
** Determined not to work - 2/14/2019
 +
** A few channels are bad - 2, 3, 4 (index, starts at 0 in software, starts at 1 on board face)
  
 
[[Category:DAQ_Testing]]
 
[[Category:DAQ_Testing]]

Revision as of 17:40, 15 February 2019

Back to Main Page >> DAQ Documentation Portal >> DAQ Testing

We will log bad boards and upload plots and descriptions of problems to: https://misportal.jlab.org/mis/apps/peer/submit.cfm

  • Run 1654 - testing a board to replace CH "vqwk1" "bad" board
    • Plots
    • It works, this board can be used - has label on top "MDBKG"
    • This board is then used to replace the board in CH "vqwk1" and Tao has taken data to establish pedestals
    • HAPLOG 3619 describes BPM 8 test with new ADC in CH crate
    • run 1233 and 1234 were used to test (run 1233 had address conflict, so ignore)
  • Run 1655 - test ADC 26
  • Run 1656 - test ADC 20 - testing original CH "vqwk1" "bad" board (top 4 channels are bad, ch's 4-7).
    • Plots
    • Top 4 channels are bad, as seen in CH crate data
  • Run 1657 - test ADC 15, the one with the strangely wired and tied on Op Amp
  • Run 1658 - test ADC 37 - ISU board that has recently had new channels go bad on it
    • Plots
    • Bad channels 2, 3, 4
    • Channels 2 and 4 are highly non-linear, basically flat lines offset from pedestal
    • Channel 3 works only in positive voltage range
    • No Japan error codes are reported
  • Run 1659 - test ADC 32 - used wrong HW address, no ACC light blinking and many VQWK read errors on telnet session
  • Run 1660-1662 - test ADC 32 - corrected HW address (needs to be 0x8400)
  • Run 1663 - test ADC 33 - ISU backup board
    • Plots
    • All but channel 6 are bad
    • No error code - just bad channel responses
  • Run 1664 - test ADC 31
    • Plots
    • There are ROC read errors for 1/3 events or so
    • All channels have crazy noise and no response to input voltage
    • Read errors get worse when plugging a signal into a channel
    • Some channels show differing response to input... characterizing problems will be a bit hard, so assume normal input stage problems? Or wiring to the later stages is bad across all channels?
  • Run 1665 - test ADC 31
    • Plots
    • Same ADC, same issues, rebooting VME crate didn't help
    • All 8 channels are bad, but some appear to have effects on their neighbors?
  • Run 1666 - test ADC 3 - a repaired board, previously tested in run 4642 in CH crate in November
  • Run 1667 - test ADC 6
    • Plots
    • All channels look good

Status

Most spare boards have problems, but Mark Taylor can fix them

Catalog

  • ADC 3
    • Fixed by Mark Taylor - 2/15/2019
  • ADC 6
    • Determined to work - 2/15/2019
  • ADC 15
    • Determined to work - 2/13/2019
  • ADC 20
    • Determined not to work - 2/13/2019
    • Channels 4-7 (top 4) do not work
    • Signal ramps to -10 V for all events, regardless of containing a voltage signal or not
  • ADC 26
    • Determined to work - 2/14/2019
  • ADC 31
    • Determined not to work - 2/15/2019
    • There are ROC read errors for 1/3 events or so
    • All channels have randomly stepping noisy signal that rails into certain regions and no clear response to input voltage
    • ROC Read errors get worse when plugging a signal into a channel
    • Some channels show differing response to input... characterizing problems will be a bit hard, so assume normal input stage problems? Or wiring to the later stages is bad across all channels?
  • ADC 32
    • Determined not to work - 2/14/2019
    • All 8 channels have interesting problems
    • It looks like regular bad amplifier in input stage problems
    • But, in the third run I took with this board it looks like the signal stopped being read by digital components entirely (run 1662, all signals == 0)
  • ADC 33
    • Determined not to work - 2/15/2019
    • Regular bad channels
    • Channel 6 "works" but probably it would be best to go ahead and give it a standard replacement treatment too?
  • ADC 37
    • Determined not to work - 2/14/2019
    • A few channels are bad - 2, 3, 4 (index, starts at 0 in software, starts at 1 on board face)